Product Description
The working principle of the reaction kettle: the reaction material is put into the single-layer glass reaction kettle, and the stirring reaction can be done by vacuuming; the external water (oil) bath can be heated by adding water or oil; and the evaporation and reflux of the reaction solution can be controlled. Constant pressure feeding and other functions. (Optional refrigeration components, with refrigeration source for low temperature reaction)
Product Parameter
Product Name | Glass Reactor | |||
Volume | 1L | 2L | 3L | 5L |
Material | GG-17 | |||
Frame Material | Stainless Steel | |||
Reaction Temp | -80-250 degree | |||
Vacuum Degree | 0.098Mpa | |||
Stirring Speed | 0-800pm | |||
Stirring Shaft diameter | 8mm | |||
Motor Power | 40W | 90W | ||
Heating Power | 1.5KW | 2KW | ||
Voltage (V/Hz) | 220V/50Hz | |||
Dimension(mm*mm*mm) | 320*330*1000 | 430*330*1200 | ||
Pot size(mm*mm) | Stainless steel 280*170mm | |||
Dimension(mm*mm*mm) | 800*480*450 | 1000*480*480 | ||
Gross Weight(KG) | 26 | 30 |
Operating Procedure
1. Check whether the power supply voltage is consistent with the specifications provided on the machine nameplate. (To ensure that the power supply is unified, please connect L-live wire N-neutral wire D-ground wire according to the wire marking on the cable at the factory.)
2. Plug in the power plug, turn on the power switch on the governor, and use the governor knob to select the appropriate speed.
3. The flow of material and the power of the motor speed may cause resonance at a certain point. Please change the speed of the motor to avoid resonance.
4. After the stirring rod is installed, it must be rotated by hand, and pay attention to whether the concentricity is good. If the concentricity is not good, loosen the heavy clamp, turn on the power after the clamp is correct, and adjust gradually from slow to fast.
5. If the sealing performance of the kettle is found to be degraded, please check the rotating bearing and oil seal in the stirring PTFE. If there is any damage, please replace it in time to ensure the vacuum degree of the equipment.
6. The glass instruments of this reaction device are all made of GG-17 low-expansion coefficient high-temperature resistant high-quality borosilicate glass. Users can prepare glass instruments such as reaction bottles. Please confirm the correct glass material to ensure safe use.
Corollary Equipment .